Loading vulnerability details…
CVSS v3.1
9.8
EPSS
0.94%
Published
Jun 29, 2020
Modified
Nov 21, 2024
Public PoC / Exploit
All weaponized →No public PoC or exploit code indexed for this CVE.
Links to public security research (Exploit-DB, Nuclei, Trickest, GitHub) for defensive use only.
Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
CVSS Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HWeaknesses (CWE)
Affected Products (3)